9-2 Accutorr 3 Operator's Manual
CAUTION
Service personnel should acquaint themselves with the test tools and make
sure that test tools and cables are applicable.
Check/Maintenance Item Recommended Frequency
Visual inspection When first installed or after reinstalled.
Power on test
1. When first installed or after reinstalled.
2. Following any maintenance or the replacement of
any main unit parts.
Battery
check
Functionality test
1. When first installed.
2. Whenever a battery is replaced.
Performance test Once a year or if the battery run time reduced significantly.
NIBP tests
Pressure check 1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of the module.
3. Once a year for NIBP tests.
4. Once every two years for SpO
2
test and Temp test.
Leakage test
SpO
2
test
Temp test
Electrical safety tests Once every two years, or as required.
9.3 Checking Monitor Information
1. Enter Maintenance mode. Refer to 3.7 Maintenance Mode.
2. Press the
hardkey to switch to system software version screen. Refer to 3.7.4
Viewing Software Version.
3. Press the
hardkey to display the version of each module.
9.4 Visual Inspection
Perform an overall inspection on the appearance of the equipment. The test is passed if
the equipment has no obvious signs of damage. Follow these guidelines when inspecting
the equipment:
Carefully inspect the case, display screen, buttons, and knob for obvious signs of
damage.
Inspect all external connections for loose connectors, bent pins or frayed cables.
Inspect all connectors on the equipment for loose connectors or bent pins.
Make sure that safety labels and data plates on the equipment are clearly legible.