Accutorr 3 Operator's Manual A-1
A Product Specifications
A.1 Classifications
The equipment is classified, according to IEC60601-1:
Type of protection against
electrical shock
CLASS I EQUIPMENT, equipment energized from an external
and internal electrical power source.
Degree of protection against
electrical shock
DEFIBRILLATION-PROOF TYPE CF APPLIED PART for SpO
2
,
NIBP, and TEMP
Mode of operation Continuous
Degree of protection against
harmful ingress of water
IPX1 (protection against vertically falling water drops)
Degree of safety of
application in the presence of
a FLAMMABLE
ANAESTHETIC MIXTURE WITH
AIR or WITH OXYGEN OR
NITROUS OXIDE
EQUIPMENT not suitable for use in the presence of a
FLAMMABLE ANAESTHETIC MIXTURE WITH AIR or WITH
OXYGEN OR NITROUS OXIDE
Degree of mobility Portable
A.2 Environmental Specifications
Main Unit
Item Operating conditions Storage conditions
Temperature (°C)
0 to 40 (without Temp module)
5 to 40 (with Temp module)
-20 to 60
Relative humidity
(noncondensing)
15% to 95% 10% to 95%
Barometric (kPa) 57.0 to 107.4 16.0 to 107.4
WARNING
The equipment may not meet the performance specifications given here if
stored or used outside the specified temperature and humidity ranges.
NOTE
The environmental specifications of parameter modules are the same as those
of the main unit.