4 Units Description
4-12
Figure 4-11 Sample/Reagent Disk Unit
A. Pull the handle to vertical direction and take out the disk assembly.
B. Remove the fan assembly by loosening five M4x8 cross pan head screws.
C. Remove the reagent refrigerating assembly by loosening seven M4x8 hexagon
socket head screws.
D. Remove the sensor assembly by loosening three M3x6 hexagon socket head
screws.
E. Remove the motor assembly by loosening four M4x12 hexagon socket head
screws.
F. Mounting the Sample/Reagent Disk Unit follows the reserve steps described as
above
Precautions:
Adjust tensile force of the synchronous belt while mounting the motor assembly.