4 Units Description
4-13
4.3.3.1 Reagent Refrigerating Assembly
Figure 4-12 Reagent Refrigerating Assembly
Figure 4-13 Refrigeration Components
The main components of reagent refrigerating assembly that need to be maintained
include refrigerating plate, temperature switch and condensing tube connector. The
steps of discounting are shown as follows.
1. Remove the refrigeration compartment assembly which is fixed to air duct by loosening
four M4x8 hexagon socket head screws.
2. Remove the water-proof glue in the circumference of condensing tube connector and
four M3x12 cross pan head screws. After replacing the condensing tube connector,

Table of Contents