Installing Your System
3-3
Nonterminal
SM&S Track
Module
100V-240V~
50/60Hz
100VA
Terminal Track
Module
100V-240V~
50/60Hz
100VA
Compressor
a.c. 110V/115V ±10%
60Hz ±2Hz
600VA
a.c. 220V/230V ±10%
50Hz ±2Hz
450VA
60Hz ±2Hz
300VA
3.2.3 General Environment
Working
Environment
Storage
Environment
Operation
Environment
Ambient
Temperature
15℃~32
-10℃~40
5℃~40
Relative Humidity
30%85%
10%90%
10%90%
Atmospheric
Pressure
70kPa106kPa
50kPa106kPa
70kPa106kPa
The environment shall be as free as possible from dust, mechanical vibrations, loud
noises, and electrical interference.
It is advisable to evaluate the electromagnetic environment prior to operation of this
system.
Do not use this system in close proximity to sources of strong electromagnetic radiation
(e.g. unshielded intentional RF sources), as these may interfere with the proper
operation.
Do not place the system near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close.
Do not place the system in direct sunlight or in front of a source of heat or drafts.
The environment shall be ventilated.
Do not place the system on a slope.
Connect only to a properly earth grounded outlet.
Only use this system indoors.
NOTE
Only store and use this system in specified environmental conditions.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray CAL 8000 and is the answer not in the manual?

Mindray CAL 8000 Specifications

General IconGeneral
BrandMindray
ModelCAL 8000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals