Operating Your System 
5-5 
 
5.5 Finishing Analysis 
 
Close the system as per the following procedure: 
 
Exit the SPU software of all analyzers in turn. The procedure is as follows: 
1) Tap "Menu">"Shutdown" on the main screen. 
 
2) Tap  "Yes"  to  perform  the  shutdown  procedure;  the  dialog  box  will  be  closed 
automatically. 
Check to see if all analyzers are ready for the probe cleanser maintenance. 
Perform  shutdown probe  cleanser  maintenance  to  each of the analyzers. See the 
operator’s manual of the hematology analyzer for details. 
Exit the CMU software. The procedure is as follows: 
1) Click the close button    on the upper right of the screen, a dialog box will 
display. 
2) Tap "OK" to close the dialog box and exit the CMU software. 
Turn off the power switch of the Auto Sample Processing System.