6 - 36 Operator’s Manual
6 Image Acquisition
2. Select [Probe] > [QSave] or press the user-defined key for “QSave” to enter the QSave dialog
box.
3. Click [Show Parameter] to view the image parameter of the current exam mode and the probe.
4. Click [Advanced] to set the following items if needed:
Set the value to TIC/TIB/TIS.
Enable the M-mark line.
The M-mark line always appears after being set when entering PW/M/TVM mode for
once. Press <PW>/<M> to enter the corresponding mode one time.
This function is unavailable for frozen dual-probe mode.
5. To save image parameters:
Click [Save] to save the current image values for the current exam mode of the certain
probe.
Click [Create] to save the current image parameters, measurements, comments, body
mark settings to the exam mode. The system will ask for a new name of the exam.
Click [Restore] to restore the probe and exam mode to factory settings.