Operator’s Manual 2 - 1
2 System Overview
2.1 Intended Use
It is intended for use in Abdomen, Gynecology, Obstetrics, Pediatric, Cardiac, Small Parts,
Musculo-skeletal, Vascular, Nerve, Urology, Intra-operative, Cephalic, Ophthalmic, Thoracic/
Pleural exams.
2.2 Contraindication
None.
2.3 Safety Classifications
• According to the type of protection against electric shock:
Class I equipment + internally powered equipment
• According to the degree of protection against electric shock:
Type-BF applied part
• According to the degree of protection against harmful ingress of water:
– The main unit is rated as IPX0
– The probes are rated as IPX7
– The foot switch (can be applied in the operating room) is rated as IPX8
– The power adapter is rated as IP22
• According to the disinfection and sterilization method(s) recommended by manufacturer:
Equipment with disinfection and sterilization method(s) recommended by manufacturer.
• According to the degree of safety of application in the presence of a FLAMMABLE
ANESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC
MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE
• According to the mode of operation:
Continuous operation
• According to the installation and use:
Portable equipment
Mobile equipment (when the system is installed on the mobile trolley)
• Does the equipment has any defibrillation-proof applied parts:
Non-defibrillation-proof applied part
• Permanently installed equipment or non-permanently installed equipment:
Non-permanently installed equipment