2 - 2 Operator’s Manual
2 System Overview
2.4 Product Specifications
NOTE:
The functions described in the operators manual may vary depending on the specific system
purchased.
2.4.1 Imaging Mode
2.4.2 Power supply
Voltage: 100 V - 240 V AC
Frequency: 50/60 Hz
Power input: 2.0-1.0A; Trolley power input: 4.4-2.5A
Battery voltage: 14.4 V
2.4.3 Environmental Conditions
Operating conditions
Ambient temperature: 0 °C ~ 40 °C
Relative humidity: 20% ~ 85% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
Storage and transportation conditions
Ambient temperature: -20 °C ~ 55 °C
B Mode B
M Mode M
C Mode Color
Power (DirPower)
D Mode PW/CW
Special imaging iScape View
Free Xros M
Free Xros CM
Tissue Doppler Imaging
TDI QA
Contrast Imaging
Contrast Imaging QA
LVO
Low MI Contrast
Strain Elastography
Stress Echo
Tissue Tracking QA
Smart 3D

Table of Contents