16 - 26 Operator’s Manual
16 Probes and Biopsy
5. Remove the needle-guided bracket:
a. Unscrew the Needle-guided bracket pinch nut, and remove the needle-guided bracket
from the probe.
b. Separate the probe and the needle-guided bracket.
NGB-007 Plastic/needle detachable needle-guided bracket
Perform the following procedure:
1. Install the needle-guided bracket:
a. Put on the sterile probe sheath.
b. Hold the probe by one hand, select the proper needle-guided bracket, and hold it with the
other hand. Align the narrow end tab of the needle-guided bracket with the groove of the
probe, then push the needle-guided bracket forward, making the tabs and the grooves of
the needle-guided bracket to match with the grooves and tabs of the probe.
1 Support of needle-guided bracket
2 Angle block
3 Guiding block
4 Groove and tab of the needle-guided bracket
5 Specification of guiding block
6 Guiding hole of the biopsy needle
7 Specification of angle block