Operator’s Manual I - 1
I EMC Guidance and 
Manufacturer’s Declaration
The system complies with the EMC standard IEC 60601-1-2: 2014.
Intended Environments: HOME HEALTHCARE ENVIRONMENT (except for near active HF 
SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for magnetic resonance 
imaging).
• The use of unapproved accessories may diminish system performance.
• Use of components, accessories, probes, and cables other than those 
specified may result in increased emission or decreased immunity of 
system.
• Operation of system, in the case that the patient physiological signal is 
lower than the minimum amplitude or value specified in the product 
specifications, results may not be obtained (results can be obtained when 
the HR is in the range of 30-250 bmp or when the QRS wave amplitude is 
between 0.5-5 mV.)
• Use of this equipment adjacent to or stacked with other equipment should 
be avoided because it could result in improper operation. If such use is 
necessary, this equipment and the other equipment should be observed to 
verify that they are operating normally.
• Use of accessories, transducers and cables other than those specified or 
provided by the manufacturer of this equipment could result in increased 
electromagnetic emissions or decreased electromagnetic immunity of this 
equipment and result in improper operation.
• Portable RF communications equipment (including peripherals such as 
antenna cables and external antennas) should be used no closer than 30 
cm (12 inches) to any part of the system, including cables specified by the 
manufacturer. Otherwise, degradation of the performance of this equipment 
could result.
• Keep a distance of at least 20 cm away from the monitor when WiFi 
function is in use.
This equipment has been tested and found to comply with the limits for a Class B digital device, 
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection 
against harmful interference in a residential installation. This equipment generates, uses and can 
radiate radio frequency energy and, if not installed and used in accordance with the instructions, 
may cause harmful interference to radio communications. However, there is no guarantee that