EasyManuals Logo
Home>Mindray>Diagnostic Equipment>ME8

Mindray ME8 Operator's Manual

Mindray ME8
312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
4 Setup
Operator’s Manual 4 - 41
If you have any questions, please contact the service engineer or your sales representative.
4.11 Security
4.11.1 Drive Encryption/Secure Data Wipe
Encrypt the patient data stored in the hard disk. The system provides two encryption methods:
Factory Default and User Define.
• Factory Default: the system is in factory state by default.
• User Define: add a user-defined password.
Perform the following procedure:
1. Select [User Define].
2. If no patient data are stored in the hard disk, click [Confirm], input the password and click
[Confirm] to finish the password setting
If the patient data are already stored in the hard disk, the system will pop-up prompts, follow
the steps below:
Setup Export Log Export the log.
Self Test Perform system self-test and restart the machine.
Recover To recover the system.
Enter Windows For password, please contact your service engineer.
Prepay Installment Display the prepay installment information.
Start/Stop WinProxy Cursor debug interface.
Touch Screen
Calibration
To calibrate the touch screen.
Needles List Replace Update the list of needles that support eSpacial Navi function.
NOTE:
The list of needles must be release version provided by
Mindray.
Needles List Recover Restore the list of needles that support eSpacial Navi function
to the last version.
Needles List Factory Restore the list of needles that support eSpacial Navi function
to the factory version.
Enable Tracking Enable the transducer and patient tracking function:
• Find transducers used with a patient by patient ID.
• Find patients who used a transducer by transducer SN.
After enabling the tracking function, you can set the maximum
number of days the data can be retained.
Fast Startup To speed up the startup process.
NOTE:
If McAfee has been installed, please disable this function
to avoid McAfee failure.
Item Description

Table of Contents

Other manuals for Mindray ME8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray ME8 and is the answer not in the manual?

Mindray ME8 Specifications

General IconGeneral
BrandMindray
ModelME8
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals