Structure and Assembly/Disassembly 9-75
9.3.37 Mother Board Assembly
The disassembly tool: cross-headed screwdriver (M3, M4), anti-electrostatic glove
1. See 9.3.16, 9.3.17, 9.3.18 and 9.3.19 and 9.3.24.1, 9.3.24.6 and 9.3.26 for details.
2. Cut 12 tie cables CHS-3X100 and one tie cable CHS-4X150 with the diagonal cutting pliers.
Unscrew the power supply terminal on the primary monitor, touch screen and control panel with
the screwdriver (M3, M4).
3. Wear the anti-electrostatic glove. Unscrew 13 M3 X 8 cross panhead screws with the
screwdriver (M3, M4) to remove the shield cover of the mother board.
The power supply
terminal on the
primary monitor,
touch screen and
control panel
M
agnetic ring