Image Optimization  5-27 
Invert 
Description  This function is used to set the display manner of spectrum. 
Operation  Tap [Invert] on the touch screen to complete the adjustment. 
Then the color bar can automatically invert when the color flow is steered to a certain 
angle to accommodate operators’ habit of distinguishing flow direction. Select [Auto 
Invert] via [Setup][System Preset][Image]. 
It is available only for linear probes. 
 
Speed 
Description  This function is used to set the scan speed of PW mode imaging. 
Operation  Tap  [Speed] on the touch screen to adjust the parameters. 
The smaller the value is, the slower the speed becomes. 
Changing the speed makes it easier to identify the cardiac cycles and to detect more 
details. 
 
T/F Res 
Description  Adjusts for a balance between time resolution and spatial resolution. 
Operation  Rotate the knob under [T/F Res] on the touch screen to adjust the parameters. 
Adjust according to the actual situation and the transducer. 
 
WF (Wall Filter) 
Description  To display the image accurately, it adjusts the cut-off used in the wall filter, and filters 
out the flow noise which is produced by vessel wall vibration. The real-time value is 
displayed in the image parameter area in the left corner of the screen. 
Rotate the knob under [WF] on the touch screen to adjust the parameters. 
Adjust the parameter according to the actual situation and the transducer. 
  Flow signals may be missing. 
 
Tint Map 
This function provides an imaging process based on color difference rather than gray 
distinction. 
  Ta p [Tint Map] on the touch screen to complete the adjustment. 
 
Gray map 
Description  Selects among post processing map curves to optimize grayscale images. 
Tap  [Gray Map] on the touch screen to complete the adjustment. 
 
Display Format 
Description  To set the display proportion of PW mode image and B mode image.