Image Optimization  5-147 
VOI Thickness 
Rotate the knob under [VOI Thickness] to adjust the thickness of VOI. 
Measurement Parameter 
Tap [Meas Parm.] to set the automatic measurement items for Rest, Valsalva, and Contraction 
frame. 
Orientation Refresh 
Tap [Refresh Ori], and the system adjusts the current image automatically to display the minimum 
levator ani muscle gap slice. 
For other parameter adjustments, refer to section ā5.11 3D/4Dā for details. 
 
 
5.21  Smart Fetal HR (Heart Rate) 
This feature is available only under Fetal Echo (High Flow), OB1 Fetal Echo, 
Fetal Echo (Low Flow) exam modes. 
 
During image acquisition, the user is recommended to stably hold the ultrasound probe 
to acquire fetal heart ultrasound video. If the Smart Fetal HR measurement is not 
satisfactory (e.g., the heart rate cycles are not marked accurately on the M-mode 
image), the user can manually modify the sampling lines of heart rate cycles on the M-
mode image or change the number of heart rate cycles. The fetal heart rate 
measurement result will be updated simultaneously. 
 
This feature is to measure the fetal heart rate automatically on M mode. 
Smart Fetal HR is an option. 
Basic procedures: 
1.  Perform B-mode scanning under Fetal Echo (High Flow), OB1 Fetal Echo, Fetal Echo (Low Flow) 
exam mode. 
2.  Tap [Smart FHR], and the button is highlighted.   
3.  Press <M> on the control panel, and the system acquires M-mode image automatically. The M 
mode image and the sampling line are displayed on the screen. 
If necessary, tap [Smart FHR Cycle] to set the number of heart rate cycle. 
4.  Press [Freeze] on the control panel. The heart rate cycles are marked on the M-mode image. And 
the fetal heart rate measurement result is displayed on the screen. 
If necessary, tap [Smart FHR Cycle] to set the number of heart rate cycle, and the measurement 
result is updated simultaneously. 
 
5.22  Smart VTI 
Smart VTI (Velocity-Time Integral) is used to calculate the CO (cardiac output) of the LVOT (left 
ventricular output tract), so as to quickly evaluate the cardiac function. 
NOTE:  1.  The cardiac package should be configured in advance. 
2.  Smart VTI is an option.