5-138  Image Optimization 
the magnetic generator and the patient for any circumstances after the freehand 3D cine is 
captured. Otherwise, the registration display for freehand image and the ultrasound image is 
affected. 
5.17.10 Parameter Settings 
The parameters of the Fusion Imaging are described below: 
Nav System 
Operation  Tap [Tracking System] to enable or disable the function. 
CT/MR/PET/freehand image and the ultrasound image build up the coordinate relation. 
The sensor indicator becomes green (navigation signal state indicator). 
 
Fusion Ratio 
 
Adjust the display ratio that the ultrasound image registers with the 
CT/MR/PET/freehand image. 
Operation  Rotate the knob beneath [Fusion Ratio] to set the ratio. 
The adjusting range is -1 ~ 1 in increments of 0.1. 
  0~1 represents that the ultrasound image registers with the CT/MR/PET/freehand 
image; the change of the fusion ratio appears on the right window side. 
The larger the value is, the better the display effect becomes for the ultrasound 
image registering with the CT/MR/PET/freehand image; and vice versa. 0 only 
relates to CT/MR/PET/freehand image on the right window side and 1 only relates 
to ultrasound image on the right window side.   
  -1~0 represents that the CT/MR/PET/freehand image registers with the ultrasound 
image; the change of the fusion ratio appears on the left window side. 
The larger the value is, the poorer the display effect becomes for the 
CT/MR/PET/freehand image registering with the ultrasound image; vice versa. -1 
only relates to CT/MR/PET/freehand image on the left window side and 0 only 
relates to ultrasound image on the left window side.   
 
Image Magnification 
 
The lesion or tumor ultrasound and CT/MR/PET/freehand image becomes clearer and 
easier to register via <Zoom> key.   
Operation  Rotate <Zoom> to zoom in or zoom out the image. 
The parameter “Z” on the right side of the image refers to the magnification value. The 
adjusting scale is: 0.8-2.0. 
 
Toggling CT/MR/PET/Freehand Anatomical Plane 
 
The Ultrasound System sets CT/MR/PET/freehand volume data to 3-dimensional 
image automatically after importing CT/MR/PET/freehand volume data. There are 
transverse plane, coronal plane, sagittal plane displaying from left to right at the 
bottom of the screen. 
It is available to toggle and register among three planes.   
Operation 
Tap [View Type] to select the CT/MR/PET/freehand anatomical planes (among the 
sagittal plane, transverse plane, coronal plane).