5-140  Image Optimization 
Reset Width (Reset W)/Reset Level (Reset L) 
Description  It initializes the window width and window level of the CT/MR/PET/freehand image. 
Operation  Tap [Reset W/L] to initialize the width/level of the CT/MR/PET/freehand image.   
 
Reset CT/MR/PET/Freehand 
Description 
It initializes CT/MR/PET/freehand image when CT/MR/PET/freehand image is cycled 
axially, panned, scrolled or zoomed, etc., after CT/MR/PET/freehand volume data is 
imported to Ultrasound System. 
Operation  Tap [Reset CT/MR] to initialize the CT/MR/PET/freehand image. 
Tint Map 
Description 
This function provides an imaging process based on color difference rather than gray 
distinction for CT/MR/PET/freehand image. 
Operations  Rotate the knob under the [Tint Map] item on the touch screen to select the map. 
The system provides 8 different color effect maps. 
Press the knob under the [Tint Map] item on the touch screen to turn on/off the 
function. 
 
Single Window 
Description  If the feature is enabled, only the image at the upper left is displayed on the image 
area. 
Operation  Tap  [Single Window]. 
 
Rotate 180° 
This function is to rotate CT/MR/PET volume data 180 degrees before registration. 
Tap [Rotate 180°]. 
 
B Display 
This function is to set whether to display B-mode image under Color fusion imaging 
mode. 
Operations  Tap [B Display]. 
If [B Display] is highlighted, the B-mode image is displayed. 
If [B Display] is not highlighted, the B-mode image is hidden. 
 
5.17.11 Rotating/Scrolling CT/MR/PET/Freehand Image 
The Ultrasound System sets the CT/MR/PET/freehand volume data to 3-dimensional image after the 
CT/MR/PET/freehand image is imported to iStation. Each of any CT/MR/PET/freehand anatomical 
planes (2-dimensional) appears on the screen via rotating or scrolling the CT/MR/PET/freehand image.