DICOM/HL7  11-15 
Name  NOTE 
Timeout 
Refers to time after which the system will stop trying to 
establish a connection to the service. Value: 5-60s, in 
increments of 5, and 15 by default. 
Listen Mode 
This function enables the ultrasound system to use the 
listen port for data receiving. 
Listen Port 
Port for ultrasound system to receive data after the listen 
mode function is activated. Here, the port should be 
consistent with that of the HL7 server port. 
For details of listen port setting, refer to settings in the 
server. 
Add  Click to add the Worklist service to the service list. 
Cancel  Click to cancel the parameter setting. 
Update 
Select an item in the service list, change the parameters 
in the above area, and click [Update] to update the item 
in the service list. 
Service list 
Delete  Click to delete the selected service in the service list. 
To set default server 
Select an item in the service list, click [Default] and you 
can see “Y” in the Default column. 
11.1.3.8  Others 
You can configure different type of DICOM services in different scenarios. 
11.1.4  DICOM Strategy Preset 
On the DICOM/HL7 preset screen, click [Set DICOM Strategy] to enter the strategy preset screen. 
 
DICOM strategy preset items are described as follows: