Probes and Biopsy 13-13
NOTE:
Observe the graph here carefully to immerse the transducer. Only soak parts of the
transducer below the strain relief.
CAUTION:
Repeated disinfection will eventually damage the transducer,
please check the transducer’s performance periodically.
Compatible Cleaner and Disinfectants
For the cleaner and disinfectants information, please refer to Mindray Transducer Disinfectant
Recommendation.
For P7-3TU, P7-3Ts and P7-3TE, refer to accompanied manual for disinfection and sterilization.
13.1.6 Environment
NOTE: For details of the P7-3TU, P7-3Ts and P7-3TE probes, refer to TEE Ultrasonic Transducer
manual.
Working condition
Use the probes in the following ambient conditions:
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
C11-3s
0℃~40 30%85%RH (no condensation) 700hPa1060hPa
SP5-1E
0℃~40 20%85%RH (no condensation) 700hPa1060hPa
SP5-1s
0℃~40 20%85%RH (no condensation) 700hPa1060hPa
SP5-1U
0℃~40 20%85%RH (no condensation) 700hPa1060hPa
C6-2Gs
0℃~40 20%85%RH (no condensation) 700hPa1060hPa
SC6-1s
0℃~40 20%85%RH (no condensation) 700hPa1060hPa
SC6-1E
0℃~40 20%85%RH (no condensation) 700hPa1060hPa
V11-3Hs
0℃~40 20%85%RH (no condensation) 700hPa1060hPa
L9-3s
0℃~40 20%85%RH (no condensation) 700hPa1060hPa
SD8-1E
18℃~30 20%85%RH (no condensation) 700hPa1060hPa