Probes and Biopsy 13-41
Needle Direction
This function adjusts the biopsy needle direction according to actual direction of
needle insertion. The iNeedle affecting region changes correspondingly.
Operation Tap [Needle Dir.] to select the direction.
B/iNeedle
This function is used to display B image and iNeedle image synchronously.
To disable or enable the function, tap [B/iNeedle] on the touch screen.
Tips: iZoom (full-screen magnification) is available on iNeedle status.
13.2.7 Removing the Needle-guided Bracket
NGB-011
Hold the probe and the needle-guided bracket, and then open the grip knob of the needle-guided
bracket.
NGB-018
(1) Screw the nut of the guiding block and remove the guiding block slightly along the direction of
the needle’s tail.
(2) Separate the residual part of the needle-guide bracket and the transducer from the needle.
(3) Screw the pinch nut of the bracket, and remove the needle-guided bracket from the transducer.