EMC Guidance and Manufacturer’s Declaration  16-5 
TABLE 4 
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY 
Resona I9 series is intended for use in the electromagnetic environment specified below. The customer 
or the user of system should assure that it is used in such an environment. 
IMMUNITY 
TEST 
TEST LEVEL 
COMPLIANCE 
LEVEL 
ELECTROMAGNETIC ENVIROMENT – · 
GUIDANCE 
Conduced RF 
IEC 61000-4-6 
3 Vrms 
0,15 MHz – 80 
3 Vrms 
0,15 MHz – 80 
Portable and mobile RF communications 
equipment should be used no closer to any 
part of system, including cables, than the 
recommended separation distance 
calculated from the equation applicable to 
the frequency of the transmitter. 
Recommended separation distance: 
6 Vrms in ISM 
and amateur 
radio bandsa 
between 0,15 
MHz and 80 
and amateur 
radio bandsa 
between 0,15 
MHz and 80 
d = 2 x
 
IEC 61000-4-3 
80MHz - 
2.7GHz 
80MHz - 
2.7GHz 
d = 1.2 x
  80 MHz  ~  800 MHz 
d = 2.3 x
  800 MHz  ~  2.7GHz 
Where, P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter manufacturer 
and d is the recommended separation 
distance in meters (m). 
Field strengths from fixed RF transmitters, 
as determined by an electromagnetic site 
survey
b
, should be less than the compliance 
level in each frequency range
c
.   
Interference may occur in the vicinity of 
equipment marked with the following 
symbol:   
 
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. 
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people.