EasyManua.ls Logo

Mindray Resona I9W - Page 393

Mindray Resona I9W
477 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EMC Guidance and Manufacturers Declaration 16-5
TABLE 4
GUIDANCE AND MINDRAY DECLARATIONELECTROMAGNETIC IMMUNITY
Resona I9 series is intended for use in the electromagnetic environment specified below. The customer
or the user of system should assure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC ENVIROMENT – ·
GUIDANCE
Conduced RF
IEC 61000-4-6
3 Vrms
0,15 MHz 80
MHz
3 Vrms
0,15 MHz 80
MHz
Portable and mobile RF communications
equipment should be used no closer to any
part of system, including cables, than the
recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
d = 1.2 x
P
6 Vrms in ISM
and amateur
radio bandsa
between 0,15
MHz and 80
MHz
6 Vrms in ISM
and amateur
radio bandsa
between 0,15
MHz and 80
MHz
d = 2 x
P
Radiated RF
IEC 61000-4-3
10 V/m
80MHz -
2.7GHz
10 V/m
80MHz -
2.7GHz
d = 1.2 x
P
80 MHz 800 MHz
d = 2.3 x
P
800 MHz 2.7GHz
Where, P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
b
, should be less than the compliance
level in each frequency range
c
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.

Table of Contents