EMC Guidance and Manufacturers Declaration 16-7
TABLE 6 Test specifications and minimum distances
Recommended separation distances between portable and mobile RF communications
equipment and system
Resona I9 series is intended for use in an electromagnetic
environment in which radiated RF
disturbances are controlled. The customer or the user of system can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and Reso
na I9 series as recommended below, according to the maximum
output power of the communications equipment. Portable and mobile radio communications
equipment (e.g. two
-way radio, cellular/ cordless telephones and similar equipment) should be used
no closer
to any part of this system, including cables, than determined according to the following
method:
Test frequency
(MHz)
Band(MHz)
Service Modulation
Maximum
power (W)
Distance
(m)
Immunity test
level (V/m)
385 380 - 390 TETRA 400
Pulse modulation
18Hz
1.8 0.3 27
450 430 -470
GMRS 460
FRS 460
FM ±5 kHz
deviation 1 kHz
sine
2 0.3 28
710
704 - 787
LTE Band
13,17
Pulse modulation
217 Hz
0.2 0.3 9
745
780
810
800 - 960
GSM 800/900,
tetra 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation
18 Hz
2 0.3 28
870
930
1720
1700 -1990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
LTE Band 1,
3,4,25,UMTS
Pulse modulation
217 Hz
2 0.3 28
1845
1970
2450 2400 -2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse modulation
217 Hz
2 0.3 28
5240
5100 -5800
WLAN,
802.11 a/n
Pulse modulation
217 Hz
0.2 0.3 9
5500
5785

Table of Contents