4-1
4 Filling and Draining
4.1 Checks before Filling
1. Check the vaporizer for damage.
2. Set the control dial to “0” position.
3. Observe use-by date for anesthetic agent.
4. Use on anesthetic delivery systems made by other manufacturers only after a functional
system check for geometry, leakage,
pressure and flow has been carried out by trained
service personnel (on each type of anesthetic delivery system).
5. After filling for the first time, wait 15 minutes for the dry wicks inside to become saturated
(The filling level of the anesthetic agent may drop. Refill if required.)
4.2 Filling the Vaporizer
CAUTION
z Take care not to spill anesthetic agent. Inhaling anesthetic agent vapor endangers
health.
NOTE
z The Anesthetic Vaporizer may only be used on the anesthetic delivery system after the
operating organization has checked all technical specifications of the Vapor and the
anesthesia system are met. Any deviations might result in incorrect concentrations
being delivered.
WARNING
z Only fill the vaporizer with the anesthetic agent specified on it.