10.2 Transport
1. Turn the control dial to “0
2. Disconnect the vaporizer from the anesthetic delivery system.
3. Drain the vaporizer completely. Then clean and disinfect the vaporizer.
4. Each vaporizer must be packed separately with care. Use original packing when possible.
If original packing is not available, use strong packing with at least 5 cm of impact-resistant
material around each vaporizer. Fasten packing securely.
WARNING
z Do not transport the vaporizer with anesthetic agent filled, or it may cause incorrect
output concentration.
NOTE
z Liquid anesthetic agents and filled vaporizers are subject to Hazardous Goods
Regulations. These regulations do not apply to the residues of anesthetic agents left in
the wick after draining.