9.2 Filling and Draining Related Faults and Remedies
Fault Cause Remedy
Anesthetic agent leaks from
the drainage outlet.
The drainage knob is not
closed.
Close the drainage knob.
Anesthetic agent leaks from
the filling system.
Seal on the filling system is
damaged.
Repair by trained service
personnel.
Anesthetic agent leaks from
overflow.
The vaporizer is filled above
the maximum mark.
Drain the vaporizer to the
maximum mark and check the
concentration.
Anesthetic agent does not flow
out when drained.
The filler cap is not opened or
the drainage outlet is blocked.
Open the filler cap or repair by
trained service personnel.
Anethetic agent does not flow
into anesthetic vaporizer by
Key Filler adapter
The inner tube is blocked by
liquid
Close the filler control.
Unplug the keyed end of
filling adapter from the
opening of the filling system.
And then let the liquid in the
inner tube draining into the
bottle.
9.3 Plug-in Adapter Related Faults and Remedies
Fault Cause Remedy
The vaporizer cannot be
disconnected.
The interlock device is still
engaged.
Disengage the interlock
device.
The plug-in adapter is not
fitted flush on anesthesia
machine plug-in connector
seals.
Engagement mechanism on
the plug-in adapter or plug-in
connector is damaged.
Excessive force used may lead
to jamming when
disconnecting the vaporizer.
Contact us immediately.
There is foreign body between
the plug-in connector and
plug-in adapter.
Remove foreign body.