4-13
8. Check that the sealing ring on the filler cap is not damaged, and screw the filler cap.
CAUTION
z Tighten the filler cap. Failure to do so may cause fresh gas and anesthetic agent to
escape when the vaporizer is switched on next time.
9. Unscrew the filling adapter.
10. Tighten the cap of the anesthetic agent bottle even if it is completely empty.
4.3 Draining the Vaporizer
WARNING
z Anesthetic agent which has been drained off must be handled, stored or disposed of as
medicament. If this is not done, there will be a risk of administering incorrect
anesthetic agents.
4.3.1 Pour Fill System
The draining steps of V60 Isoflurane Pour Fill Vaporizer and V60 Sevoflurane Pour Fill Vaporizer are
the same.
CAUTION
z Take care not to spill anesthetic agent. Do not inhale anesthetic agent vapor which
endangers health.
NOTE
z Do not drain liquid anesthetic agent into an open container or significant quantities of
anesthetic agent vapor will escape.
z Do not reuse the anesthetic agent drained from the vaporizer.