4-14
Place the vaporizer upright or suspend so that all the anesthetic agent can drain out.
1. Turn the control dial clockwise back to “0” position until the “0” button pops up.
2. Hold the correct bottle for the anesthetic agent being drained below the drainage outlet at the
bottom of the vaporizer.
3. Unscrew the filler cap counter clockwise slowly.