A-2
A.3 Operating Range
Temperature
During operation 10 to 40
During storage (empty) -20 to 60
During transport (empty) -20 to 60
Humidity
During operation 15 to 95%, non-condensing
During storage 10 to 95%, non-condensing
Atmospheric pressure
During operation and shut-down (filled,
control dial at “0” position)
70 to 106 kPa
During storage (empty) 50 to 120 kPa
Concentration range
Isoflurane 0 to 6%
Sevoflurane 0 to 8%
Enflurane 0 to 5%
Halothane 0 to 5%
A.4 Performance Specifications
Range of concentration accuracy (maximum value always applies)
Operating environment
15 to 35 or 0.2 to 10 L/min 10 to 15 or 35 to 40 or 10 to
15 L/min
Set concentration 6% ±0.20 vol.% or ±20% rel.,
whichever is greater
+0.30/-0.20 vol.% or +25/-20%
rel., whichever is greater
Set concentration > 6% ±0.25 vol.% or ±20% rel.,
whichever is greater
+0.35/-0.25 vol.% or +30/-20%
rel., whichever is greater
Maximum angle of tilt
Alone, freestanding 10º
During operation (fixed position) 30º
Pressure difference
Difference between pressure range and ambient pressure on the
vaporizer outlet
-10 to 10 kPa