4-5
4. Check the filling level on the sight glass during filling. The vaporizer must be hanged
vertical or stand upright during this check.
During the filling process, the filling level must not exceed the maximum mark, or there is a
risk of incorrect output concentration. If the maximum mark has been exceeded, the agent
will flow out .Please drain the excess liquid (see 4.3 Draining the Vaporizer) until the level
drops below the maximum mark.
5. When the maximum mark is reached, stop pouring agent. If the vaporizer is filled above the
maximum mark by a few millimeters, the anesthetic agent will start to overflow through the
overflow hole.
CAUTION
z When the anesthetic vaporizer is overflowed, the anesthetic agent will escape to the
environment.
6. Tighten the filler cap clockwise. If this is not done properly, fresh gas and anesthetic agent
may escape when the vaporizer is switched on next time.
7. Tighten the cap of the anesthetic agent bottle even if it is completely empty.
Minimum mark
Maximum mark