4-10
8. Open the filler control and the liquid agent will flow into the vaporizer.
9. Check the filling level in sight glass during filling. When the maximum mark is reached,
flow stops automatically.
CAUTION
z If the connection between the filling adapter and anesthetic agent bottle or that
between the filling adapter and filling system is not leak-tight, anesthetic agent may
continue to flow into the vaporizer.
10. Close the filler control.
11. Put down the anesthetic agent bottle slowly.
12. Unscrew the lock knob.
13. Pull the keyed end of the filling adapter out of the filling system.
14. Put the keyed block back into the opening of the filling system.
15. Tighten the lock knob.
16. Unscrew the filling adapter.
17. Tighten the cap of the anesthetic agent bottle even if it is completely empty.
CAUTION
z If the connection between the keyed block and filling system is not leak-tight, anesthetic
agent may escape, endangering health.
Maximum mark
Minimum mark
O
en the filler control