4-17
6. Remove the keyed block.
7. Push the keyed end of the filling adapter into the opening of the filling system until it cannot
move.
8. Tighten the lock knob clockwise.
CAUTION
z If the connection between the filling adapter and filling system is not leak-tight,
anesthetic agent may escape.