6-2
2. The sealing rings on the anesthesia machine plug-in connector must be undamaged. There
should be no foreign bodies on the plug-in connector.
3. Switch the vaporizers off when one or more than one vaporizers have been on the manifold
of the anesthetic delivery system, before hanging on another vaporizer.
4. Set the control dial to “0” position.
5. Hold the vaporizer in vertical position with both hands and lower gently onto the anesthesia
machine
plug-in connector.
The sealing rings
are undamaged.