13
All parts of the breathing system can be cleaned and disinfected. The cleaning
and disinfection methods are different for different parts. You need to select
the appropriate method to clean and disinfect the parts based on the actual
situations to avoid cross-contamination.
This table is our recommended cleaning and disinfection methods for all parts of
the breathing system.
Parts
Intermediate level
disinfection
High level
disinfection
A* B* C*
Breathing tubes and Y piece
★★
Breathing mask
★★
Flow sensor
Bellows assembly
★★
Inspiratory and expiratory check valves
assembly
★★
O2 sensor
Canister assembly
★★
Canister connection block assembly
★★
Water collection cup
★★
Bag arm
★★
BYPASS assembly
★★
Breathing System
★★
Manual bag
★★
AGSS assembly
Indicates that this disinfection method is applicable.
A* Clean with a damp cloth soaked in mild detergent and then wipe off the
remaining detergent with a dry lint free cloth.
B* Flush with water rst; then soak in water and cleaning solution (water
temperature 40ºC recommended) for approximately three minutes and wipe
with 70% ethanol.
C* Steam autoclave at maximum 134ºC.
Cleaning and Disinfection

Other manuals for Mindray WATO EX-65

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray WATO EX-65 and is the answer not in the manual?

Mindray WATO EX-65 Specifications

General IconGeneral
TypeAnesthesia Machine
ModelWATO EX-65
BrandMindray
Tidal Volume Range20 to 1500 mL
O2 Concentration21% to 100%
I:E Ratio4:1 to 1:8
Ventilation ModesVCV, PCV, SIMV, Manual, Spontaneous
Flow MeterO2, N2O
VaporizerUp to 2 vaporizers
PEEP0 to 20 cmH2O
Gas SupplyO2, N2O, Air
MonitoringAirway Pressure, Flow, Volume
O2 Flush35 to 75 L/min
Respiratory Rate4 to 100 breaths per minute
Power Supply100 to 240 VAC, 50/60 Hz
Fresh Gas Flow0 to 15 L/min