14
Clean and Disinfect the Anesthesia Machine Housing
Clean the surface of the anesthesia machine housing with a damp cloth soaked in
mild detergent (such as 70% ethanol).
After cleaning the housing, remove the remaining detergent by wiping with a dry lint
free cloth.
1.
2.
WARNING
Seeping liquid into the control assembly can damage the equipment or cause
personal injury. When cleaning the housing, make sure that no liquid ows into
the control assembly and always disconnect the equipment from the AC mains.
Reconnect the AC mains after the cleaned parts are fully dry.
NOTE
Use only soft dry and lint free cloth to clean the display. Do not use any liquid
for display cleaning.
Disassemble the Breathing System Cleanable Parts
You need to disassemble the breathing system cleanable parts first before
cleaning the system.
O2 Sensor
1.Remove one end of the O2 sensor cable
from the connector on the anesthesia
machine. Unplug the other end of the cable
from the O2 sensor.
2.
Turn the O2 sensor counterclockwise to
take it out.
Cleaning and Disinfection

Other manuals for Mindray WATO EX-65

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray WATO EX-65 and is the answer not in the manual?

Mindray WATO EX-65 Specifications

General IconGeneral
TypeAnesthesia Machine
ModelWATO EX-65
BrandMindray
Tidal Volume Range20 to 1500 mL
O2 Concentration21% to 100%
I:E Ratio4:1 to 1:8
Ventilation ModesVCV, PCV, SIMV, Manual, Spontaneous
Flow MeterO2, N2O
VaporizerUp to 2 vaporizers
PEEP0 to 20 cmH2O
Gas SupplyO2, N2O, Air
MonitoringAirway Pressure, Flow, Volume
O2 Flush35 to 75 L/min
Respiratory Rate4 to 100 breaths per minute
Power Supply100 to 240 VAC, 50/60 Hz
Fresh Gas Flow0 to 15 L/min