ZS3 Instructions for Use
70 | P a g e
Image Review Options
Images can be reviewed during an active exam or after an exam has been ended.
Onscreen Thumbnails: Point to and click on onscreen thumbnails to view, edit, send to
media, Tag, or delete the selected image or clip.
Current Exam: Press Keyboard control to display images and clips from the currently
active exam.
Archive: Press Keyboard control to display a list of previous exams. By default this
shows the list of exams present on the internal hard drive.
A drop-down menu on this screen (Data Source) will switch to a list from
media inserted in the USB port.
While a study is highlighted (from Archive or Current Exam), pointing to a thumbnail
along the bottom displays the following tools, which indicate that individual images
can be deleted, transferred to media, or Tagged etc. independent of the entire
collection.
Delete , Send To , and Tag options
New Patient: This displays the patient demographic screen for new patients. This
screen also
Has a Restart button to restart previously completed exams.
Has a Pause button to pause the exam.
Has a Resume button to resume a previously paused exam.
Has an Archive button which provides an alternate means of accessing the
Archive.
Total Recall (TRI) Options
The Total Recall Imaging (TRI) feature allows the user to store pre-scan converted
ultrasound data; review/manipulate that data during an active exam or after an exam has
been ended; and re-save the manipulated data.
The onscreen thumbnail shall include an ICON reflecting the availability
of Total Recall data as indicated by TRI.
When the Total Recall data has been loaded, the ICON onscreen will
be updated with a checkmark.
Total Recall Review Methods: If Total Recall Data has been stored with an image or clip, the
user can:
Automatically Load Total Recall Data: Select the option ‘automatically load Total
Recall Data when images selected in Review or Archive’ on the Setup/ General tab.
When this option has been selected, Total Recall data will automatically be
loaded when images or clips are selected for Review
An OLED will then become available to Exit TRI’ or ‘Delete TRI’.