EasyManua.ls Logo

Misonix neXus - Footswitch Functionality and Pairing

Default Icon
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions For Use neXus Ultrasonic Surgical Aspirator
Doc #: 100-10-1000 rev A p 45
8.3 Power Supply Faults
The console monitors the internal power supplies at all times and faults in cases when they are compromised. A System
Reset screen is displayed together with an audible indicator. Ultrasound and Irrigation are deactivated.
Fault Type
Fault Screen
Fault Action
System Reset
Power Supply Fault
Example Screen
System functions automatically
disabled.
Pressing the System Reset Button
reboots the system.
“neXus” splash screen appears
followed by the Priming screen.
Priming screen may be skipped.
*Skipping priming screen returns
user to the Main screen.
Possible Cause
Corrective Action
Damaged console
Press the System Reset Button and try again.
If corrective action steps above are followed and fault continues, the console
may need to be replaced.
Table 8.3: Power Supply Faults
8.4 Communication Faults
The console monitors the internal microprocessors at all times and faults in cases when they are not communicating
properly. A System Reset screen is displayed together with an audible indicator. Ultrasound and Irrigation are deactivated.
Fault Type
Fault Screen
Fault Action
System Reset
Communication Fault
Example Screen
System functions automatically
disabled.
Pressing the System Reset Button
reboots the system.
“neXus” splash screen appears
followed by the Priming screen.
Priming screen may be skipped.
*Skipping priming screen returns
user to the Main screen.
Possible Cause
Corrective Action
Damaged console
Press the System Reset Button and try again.
If corrective action steps above are followed and fault continues, the console
may need to be replaced.
Table 8.4: Communication Faults

Related product manuals