Service, Repair and Technical Correspondence
WARNING Proper system grounding can only be ensured when an approved, hospital-grade receptacle and matching
power cord are used. To avoid the risk of electric shock, this equipment must only be connected to a supply
with protective earth. Install plug and receptacles as per local regulations before operating the unit. Power
cord, plug and receptacle should be examined to verify that they are in good working condition before
connecting the console. Never pull on the power cord to remove it from the receptacle.
WARNING
The neXus console automatically adjusts for the mains voltage and frequency. Confirm that the correct
fuses are being used. Refer to section 10 in instructions for fuse replacement
.
WARNING The NeXus® Ultrasonic Surgical Aspirator System generates high voltages within the console itself and the
connected handpiece. To avoid injury, the console should never be operated before ensuring that its cover
is properly closed and not tampered with. Do not attempt to remove or disassemble the cover. There are
no user-serviceable parts inside the console. All service should only be performed by an authorized Misonix
representative. No modification of this equipment is required
10.1 Fuse Replacement
CAUTION The only user replaceable fuses are the two fuses located on the bottom rear of the unit. Replacement fuses must be
identical in type, voltage rating and current rating to the original fuse.
WARNING Replacement fuses other than what is specified can cause a fire hazard. Use only as specified.
WARNING If a Mains Power fuse fails after replacement when the unit is reactivated, discontinue use of the
device and contact an authorized Misonix representative.
Table 10.1 Console Fuse Specifications