3. OPERATING PROCEDURE
No. 99MAH016B
3-3
Switching the analog range (analog resolution) during display of 3.2.3
range
1 Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “range” on the LCD starts blinking.
2 Quickly pressing the [SET/ZERO] key changes range from blinking to lighting-up
and blinks the currently-set analog range.
3 Press the [MODE] key quickly to switch the setting to a desired analog range.
Digital
resolution
Selectable analog range
0.0005mm
±0.01
±0.02
±0.05
±0.1
±0.2
±40
±80
0.001mm
±0.02
±0.04
±0.1
±0.2
±0.4
±40
±80
.00002”
±.0004
±.0008
±.002
±.004
±.008
±2
±4
.00005”
±.001
±.002
±.004
±.01
±.02
±2
±4
.0001”
±.002
±.004
±.01
±.02
±.04
±2
±4
RS-232
command *
A01
A02
A05
A10
A20
AFS
AFL
*: For information about RS-232 commands, refer to section 4.2.4.
4 Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “dir”.
Switching the count direction during display of “dir” 3.2.4
1 Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “dir” on the LCD starts blinking.
2 Quickly pressing the [SET/ZERO] key changes “dir” from blinking to lighting-up. Only
“dir” is displayed if the currently-set direction is “+” count, and “REV” blinks at the left
of “dir” if the direction is “-" count.
3 Press the [MODE] key quickly to switch the setting to a desired direction (between +
and -).
4 Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “output”.
NOTE
Upon change of the count direction, the sign (+/-) of a display value is also changed.
Before calling a preset value, perform setup of the count direction.
Switching the I/O format between SPC(digimatic) and RS-232 during 3.2.5
display of “output”
1 Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “output” on the LCD starts blinking.
2 Quickly pressing the [SET/ZERO] key changes “output” from blinking to lighting-up
and blinks the currently-set I/O format.
3 Press the [MODE] key quickly to switch the setting to a desired I/O format (between
SPC and RS-232).
4 Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “id” (section 3.2.6) if SPC is specified. If RS-232 is specified as the I/O format,
the procedure is continuously switched to the RS-232 setup (section 3.2.5.1).
RS-232 setup 3.2.5.1
1 When the procedure is switched to the RS-232 setup, “bps” starts blinking.
2 Press the [SET/ZERO] key quickly to blink the currently-set baud rate.
3 Press the [MODE] key quickly to switch the setting to a desired baud rate (4800 or
9600).
4 Quickly press the [SET/ZERO] key to apply the communication setting and switch the
setup item to the next “p-bit.”.