No. 99MAH016B
procedure in steps 2 to 4.
6 When the data bit setting has been applied, the setup item is switched to the next “id”.
Initial setting (factory defaults) for RS-232 is as follows:
Bit rate: 9600bps Parity bit: even Data bit: 7bit
Setting ID No. for RS-232 communication and remote control during 3.2.6
display of “id”
1 Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “id” on the LCD starts blinking.
2 Quickly pressing the [SET/ZERO] key changes “id” from blinking to lighting-up and
blinks the currently-set ten’s place of ID No.
3 Press the [MODE] key quickly to switch the ten’s place to a desired number (0 to 9).
4 Press the [SET/ZERO] key quickly to apply the specified number and blink the current
unit’s place of ID No.
5 Press the [MODE] key quickly to switch the unit’s place to a desired number (0 to 9).
4 Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “start”.
When the ID No. of the instrument has been set, only the signals from the remote
control and RS-232 communication with the same ID No. are received. If ID No. on the
remote control and RS-232 communication sides is set to “00”, however, the
instrument receives any signals irrespective of its ID No.
The remote control (option) ID No. can be set to one of “00” to “07” or “09” to “15”.
The RS-232 communication ID No. can be set to one of “00” to “09”.
Switching the length measuring system during display of “start” 3.2.7
1 Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “start” on the LCD starts blinking.
2 Quickly pressing the [SET/ZERO] key changes “start” from blinking to lighting-up and
blinks the currently-set length measuring system.
3 Press the [MODE] key quickly to switch the setting to a desired length measuring
system.
Always starts up under the INC system.
Starts up under the length measuring system at the most
recent shutdown.
4 Quickly press the [SET/ZERO] key to apply the setting and switch the setup item to
the next “reset
Returning all parameters to factory defaults during display of “reset” 3.2.8
1 Put the instrument in the Parameter Setup mode, and then quickly press and release
the [MODE] key repeatedly until “reset” on the LCD starts blinking.
2 When pressing the [SET/ZERO] key quickly, “no” on the LCD starts blinking.
3 To return the parameters to defaults, press the [MODE] key quickly to switch to “yes”
under blinking.