4.2.4 Communication command (I/O format)
1 Measurement command
Operation content
Input
(External device -> this instrument)
Output
(This instrument -> external device)
Measurement
system
Zero-setting
(Moves to the INC
system)
CR**CRLF
CH**CRLF
Moves to the Preset
system
DS**,PCRLF
DH**,PRESETCRLF
*Refer to the NOTE below
DH**CRLF
Preset value setting
CP**,+0016.2345CRLF
CH**CRLF
Preset value output
DP**,OUTCRLF
DH**,+0016.2345CRLF
Measurement
mode
switching
Normal
CN**CRLF
CH**CRLF
Max.
CX**CRLF
Min.
CM**CRLF
TIR (Run-out)
CW**CRLF
Data output
Normal value
GA**CRLF
GN**,+0016.2345CRLF
Max. value
GX**,+0016.2345CRLF
Min. value
GM**,-0016.2345CRLF
TIR (run-out) value
GW**,+0016.2345CRLF
Peak hold
Cancel
(Peak hold start)
CL**CRLF
CH**CRLF
Peak zero set
DS**,XM-ZEROCRLF
DH**CRLF
Tolerance
judgment
Judgment ON
DJ**,ONCRLF
DH**CRLF
Judgment cancel
DJ**,OFFCRLF
Upper limit value input
CG**,+0016.2345CRLF
CH**CRLF
Lower limit value input
CD**,-0016.2345CRLF
Upper limit value output
DJ**,GOUTCRLF
DG**,+0016.2345CRLF
Lower limit value
output
DJ**,DOUTCRLF
DD**,-0016.2345CRLF
Result output (OK)
DJ**,OUTCRLF
DH**,OKCRLF
Result output (+NG)
DH**,+NGCRLF
Result output (-NG)
DH**,-NGCRLF
Status output
(under cancel)
DH**,JOFFCRLF
Analog display
Centering
DA**CRLF
DH**CRLF
Function lock
Lock
DF**,LOCKCRLF
DH**CRLF
Cancel
DF**,FREECRLF
Status output
(under lock)
DF**,OUTCRLF
DH**,F-LOCKCRLF
Status output
(under cancel)
DH**,F-FREECRLF
NOTE
For the detail of each item in the operation content, refer to Chapter 3.
“CRLF” represents “CR” (carriage return) and “LF” (line feed).
** indicates an ID No. For information about ID No. setup method of this
instrument, refer to section 3.2.
Set an ID No. between 00 and 99.
Digimatic indicators with a specified ID No. between 01 and 99 can receive and execute a
command.