No. 99MAH016B
4-6
IMPORTANT
For the Initial setting by RS-232 input, setting of both Input/output format” and ID
No. are retained.
NOTE
For the detail of each item in the operation content, refer to Chapter 3.
A## indicates the setting command of the analog range. For information about the
analog range, refer to section 3.2.3. Switching the analog range (analog resolution)
during display of range
3 Error command
Operation content
Input
(External device -> this instrument)
Output
(This instrument -> external device)
Error
Over-speed
CH**,Error 20CRLF
Overflow
CH**,Error 30CRLF
Communication
command error
CH**,Error 52CRLF
Tolerance setup
error
CH**,Error 90CRLF
Preset value error
CH**,Error 95PCRLF
Upper limit
value error
CH**,Error 95GCRLF
Lower limit
value error
CH**,Error 95DCRLF
Cancel
CS**CRLF
CH**CRLF
NOTE
indicates a space.
For detailed information about errors, refer to “5 ERROR MASSAGE AND ACTION”.
4.2.5 Communication sample
The example of RS-232 communication by using the hyper terminal (attached in the
Windows) is described below.
1 Start up the hyper terminal and select the setting of the digimatic indicator (ID-H side)
to the Port setting.
It is also possible to select Port setting by clicking on File Properties
Connect to and press the Configure…” button.
When it is necessary to set the initial setting, set Port setting as follows.
Bits per second9600bps, Data bit7bit, Parity even
Stop bits2bit, Flow controlHardware
2 Continue to select some of the settings for the hyper terminal.
Click on File
Properties
Settings to change the setting as follows.
Emulation」:ANSI
Click on ASCII setup to put the checking mark for check boxes: Send line ends
with line feeds, Echo typed characters locally, Append line feeds to incoming line
ends and Wrap lines that exceed terminal width.
3 Example of communication
The procedure below is the case when the tolerance judgment is performed after presetting