4. DATA INPUT/OUTPUT
No. 99MAH016B
4-7
+12.3455mm and setting the upper limit for 12.35mm and the lower limit 12.34mm in the
preset system. For this time, ID No. is 00 (default).
(1) Perform
presetting: Input the command CP00,+0012.3455 from PC. This
instrument performs presetting and responds the command CH00.
(2)
Set Upper limit: Input the command CG00,+0012.3500 from PC. This instrument
sets the upper limit and responds the command CH00.
(3) Set
Lower limit: Input the command CD00,+0012.3400 from PC. This instrument
sets the lower limit and responds the command CH00.
(4)
Set the tolerance judgment mode. Input the command DJ00,ON from PC. This
instrument changes the measurement mode to the tolerance judgment mode and
responds the command DH00.
TIP
For the example of RS-232 communication described above, perform presetting (change of
the measurement system) at the beginning. This instrument changes the upper and the
lower limit of the measurement mode when inputting the command.
For the procedure (2) to (4), it is allowed to proceed with the random order.
When using the hyper terminal, it is not possible to correct the command typing error. Once
input the incorrect command (this instrument responds the command CH**,Error 52”)
and then input the correct command.