Guidance and Manufacturers Declaration – Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specied below.
The customer or the user of this device should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment – Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
ISM
(c)
frequency band:
6 V
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
27 V/m
385 MHz
28 V/m
450 MHz
9 V/m
710, 745, 780 MHz
28 V/m
810, 870, 930, MHz
28 V/m
1720, 1845, 1970 MHz
28 V/m
2450 MHz
9 V/m
5240, 5500, 5785 MHz
3 V
ISM
(c)
frequency band:
6 V
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
27 V/m
385 MHz
28 V/m
450 MHz
9 V/m
710, 745, 780 MHz
28 V/m
810, 870, 930, MHz
28 V/m
1720, 1845, 1970 MHz
28 V/m
2450 MHz
9 V/m
5240, 5500, 5785 MHz
Portable and mobile RF communications equipment
should be used no closer to any part of this device,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distances
d
= 1.2 150 kHz to 80 MHz
d
= 1.2 80 MHz to 800 MHz
d
= 2.3 800 MHz to 2.7 GHz
d
= Portable wireless RF communication
equipment
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer,
E
is the compliance level in V/m
and
d
is the recommended separation distance in
meters (m).
Field strengths from eld RF transmitters, as
determined by an electromagnetic site survey
(a)
,
should be less than the compliance level in each
frequency range
(b)
.
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reection from structures, objects and people.
(a) Field strengths from xed transmitters, such as base stations for ratio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicated theoretically with accuracy. To assess
the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be considered. If the
measured eld strength in the location in which this device is used exceeds the applicable RF compliance level above, this
device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting of relocating this device.
(b) Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.
(c) The ISM (Industrial, Scientic and Medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553
MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Morita Veraview iX and is the answer not in the manual?

Morita Veraview iX Specifications

General IconGeneral
BrandMorita
ModelVeraview iX
CategoryDental equipment
LanguageEnglish

Related product manuals