EasyManua.ls Logo

MSA BD mini - Maintenance and Care; Maintenance Instructions; Maintenance Intervals

MSA BD mini
204 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MSA MAINTENANCE AND CARE OF THE SCBA
BD mini 19
GB
4. Maintenance and Care of the SCBA
4.1. Maintenance Instructions
This product should be regularly checked and serviced by specialists. Inspection
and service records must be maintained. Always use original parts from MSA.
Repairs and maintenance must be carried out only by authorised service centres
or by MSA. Changes to devices or components are not permitted and could result
in loss of approved status.
MSA is liable only for maintenance and repairs carried out by MSA.
Do not use organic solvents like alcohol, white spirit, petrol, etc.
When drying/washing, do not exceed the maximum permissible temperature of
60°C.
4.2. Maintenance Intervals
MSA recommends the following maintenance intervals. If needed and
by considering the usage, tasks may be at even shorter intervals than
indicated.
Observe national laws and regulations!
If in any doubt, ask your local MSA contact person.
Component Work to be
Performed
Before
use
After use Annu-
ally
Every 3
years
Every 9
years
1)
Cleaning X X
Sight, function and
tightness check
X X
Compressed air
breathing apparatus
complete
Check by user
2)
X
Compressed air
breathing apparatus
without cylinder and
lung governed
demand valve
Overhaul X
Filling pressure
check
X
Compressed air
cylinder with valve
Technical expert
test
See instruction manual for compressed air
cylinder. Please observe national rules!
Lung governed
demand valve
See instruction manuals for lung governed demand valve / full face
mask. Please observe national regulations!
3)
1)
For SCBA apparatus that are frequently used, we recommend a complete overhaul after
approx. 540 hours. For example, this corresponds to 1080 applications with a duration of 30 minutes.
2)
The checks are performed with the respective lung governed demand valves and if required, with the
respective full face masks.
3)
Rubber components are subject to ageing with varying rates and according to local conditions and
must be checked and replaced at regular intervals.