16 | ni.com | NI PXIe-4302/4303 and TB-4302/4302C
Keep the ambient temperature as stable as possible.
Keep the terminal block, module, and chassis in a stable and consistent orientation.
Install the PXIe filler panels in any unused slots.
When stacking the system, avoid stacking the chassis above a heat source as heat rises and
can introduce error.
Allow the thermal gradients to settle after a change in system power or in ambient
temperature. A change in system power can occur when the system powers on. Refer to the
warm-up time in the NI PXIe-4302/4303 Specifications for more information.
If your PXIe chassis has fan filters, clean or replace them regularly to ensure proper system
cooling. Refer to the PXI/PXI Express Chassis Fan Filter Kit document for more
information.
Note Refer to the Maintain Forced-Air Cooling Note to Users for more
information about maintaining optimal forced-air cooling for PXI/PXI Express
devices.
More Information
After you install NI-DAQmx, the NI-DAQmx documentation is available from Start»All
Programs»National Instruments»NI-DAQmx. Additional resources are online at
ni.com/
gettingstarted
.
You can access online device documentation by right-clicking your module in MAX and
selecting Help»Online Device Documentation. A browser window opens to
ni.com/
manuals
with the results of a search for relevant documents. If you do not have Web access,
documents for supported modules are included on the NI-DAQmx disc.
Troubleshooting
Go to ni.com/support/install or ni.com/kb.
If you need to return your National Instruments hardware for repair or device calibration,
go to
ni.com/info and enter rdsenn to start the Return Merchandise Authorization
(RMA) process.
Specifications (TB-4302/4302C)
Note NI PXIe-4302/4303 module specifications are located in the
NI PXIe-4302/4303 Specifications document.
All performance specifications are typical unless otherwise noted. These specifications are valid
within the full operating temperature range.