Instructions for Use XLTEK Protektor32
74
Multiple Connections
You can monitor more than one Acquisition Station by more than one Remote Station. The following
configurations below are the available:
Both Local Portal (LP) and Remote Portal (RP) can run on the same station at the same time.
RP can connect to several acquisition stations (nLP) and monitor them at the same time (one-to-
many monitoring).
LP can accept several observers (nRP) at the same time (many-to-one monitoring).
Above three points give many-to-many monitoring where all topologies (including loops) are allowed.
LP-nRP textual chat.
Additional Information and Terms
Bandwidth – is the total amount of EPWorks data being sent to the observer (the units are in
bytes/second
Messages refers to how many messages per second are being sent to the observer (data is
passed in messages)
Queue refers to how many messages are currently in the process of going across to the observer
Remote monitoring Optimizations These are all optimizations that can reduce system load and
network load both on the acquisition system and the remote system
SSEP by default, not every single “intermediate” sweep is send of an averaged waveform in
order to save bandwidth. The user can specify how frequently the intermediate sweeps are
sent (final completed sweeps are always sent).
Update free run EMG every n seconds this saves network bandwidth by sending sweeps
every few seconds. The user can specify 0 and it will send the data as soon as they have it.
Accumulated EEG this specifies how “smoothly” the EEG on the remote system will be
displayed. The more frequently it is sent, the more the network bandwidth it will use.
Remote Monitoring Physician will be able to add notes directly into the patient record “log book”
while connected remotely.
Connectivity a message will appear on both the acquisition system and the remote monitoring
system if a network connection is lost and also when the connection is reestablished..

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the natus XLTEK Protektor32 and is the answer not in the manual?

natus XLTEK Protektor32 Specifications

General IconGeneral
Brandnatus
ModelXLTEK Protektor32
CategoryMedical Equipment
LanguageEnglish

Related product manuals