Oximeter Configurable Settings
The N-200 provides the operator with the capability to tailor
the system for specific clinical applications. Capabilities
include:
• Audible alarms that can be silenced; the alarm has
adjustable volume.
• C-LOCK ECG synchronization that enhances oximetry
signal processing during patient movement or for patients
with low perfusion.
• Three oximetry operating modes that change measurement
averaging time to suit varied clinical applications.
• Oximetry trend memory, with up to 12-hour SpO2 and
pulse rate trend data storage.
• Oximetry and pulse rate event memory, with 1-hour event
data storage. Data storage of event memory markers
includes: alarm-limit-defined events and user-defined
events.
• Analog and digital output of saturation, pulse rate, and
pulse waveform data. When an ECG signal is provided to
the patient module three-lead ECG connector, the N-200
provides an analog output of the ECG waveform.
WARNING: Do not use the ECG analog output as a
trigger for synchronous defibrillation.