Specifications
89
Sensor
Within physiologic range 82.4–107.6°F (28–42°C) for accurate
measurement.
Humidity
Any humidity/temperature combination without condensation.
Altitude
0–10,000 ft (0–3,048 m)
ELECTRICAL CHARACTERISTICS
Protective Class
Class I: mains-supplied unit using a protective ground
Degree of Protection
Type BF: patient electrically isolated
Voltage
100 VAC -10% to 120 VAC +10%, 50/60 Hz
Power Consumption
Maximum rating: 25 VA
Mains Fuse Rating
1 x T 0.5 A, slow-blow, 250 V
Leakage Current
50 µA maximum, power line to ground
10 µA maximum, patient connector to ground
10 µA maximum, patient connector to power line