NELLCOR Sensors
24
Table 2: Selected NELLCOR Sensors
Oxygen Transducer
Model
Patient
Weight
OXISENSOR
¨
II or OXISENSOR
¨
(sterile, single use)
N-25
I-20
D-20
D-25(L)
R-15
<3 kg, >40 kg
3 to 20 kg
10 to 50 kg
>30 kg
>50 kg
OXICLIQ
¨
(sterile, single use) A
P
>30 kg
10 to 50 kg
DURA-Y
¨
(nonsterile, reusable) D-YS >1 kg
OXIBAND
¨
(nonsterile, reusable) OXI-A/N
OXI-P/I
<3 kg, >40 kg
3 to 40 kg
DURASENSOR
¨
(nonsterile,
reusable)
DS-100A >40 kg
NELLCOR reflectance sensor
(nonsterile, limited reuse)
RS-10 >40 kg
CLEANING AND REUSE
Caution: Do not sterilize any NELLCORsensor or
adhesive by irradiation, steam, or ethylene oxide, and
do not immerse in water or cleaning solutions.
Clean the surface of reusable sensors by wiping with an agent
such as 70% isopropyl alcohol. When using adhesive sensors,
use a new adhesive sensor, wrap, or bandage for each patient.