Connecting to Other Instruments
66
CONNECTING TO THE N-7500 NETWORK
Refer to the N-7500 operator’s manual for full instructions.
Nellcor’s OXINET N-7500 pulse oximetry network is a radio-
frequency telemetry system that displays data from as many as
eight NELLCOR pulse oximeters, such as the N-200, on its
central monitoring station. The installed N-7500 includes radio
interfaces, connected NELLCOR pulse oximeters, and the
central monitoring station.
To connect the N-7500 to the N-200, set the rear-panel DIP
switches as follows: switches 3, 4, 5, and 7 down; switch 8 up.
CONNECTING THE N-9000 RECORDER/INTERFACE
Connect the N-9000 to the N-200 as follows:
1. Set the N-200 baud rate switches to 2400 (switch 7 down; 8
up).
2. Set the N-200 RS-232 FORMAT switches to the N-7000/
N-9000 recorder format (switch sections 3–5 down).
3. Connect the gray end of the fiber-optic cable provided with
the N-9000 to the DATA connector on the N-200
powerbase. Connect the blue end to the N-9000 connector
labeled DATA INPUT. Push each connector in until it
snaps into place. Route the fiber-optic cable without
bending it; poor transmission may occur if the cable is
kinked.
Refer to the N-9000 operator’s manual for complete operating
instructions.